Material: | Medical Grade PVC | Length: | 5cm 8cm 10cm 13cm 15cm 16cm 20cm 25cm |
---|---|---|---|
Catheter Size: | 6.5Fr 7Fr 7.5Fr 8Fr 10Fr 11.5Fr 12Fr | Lumen: | Single/Double/Triple |
Material Of Extension Tube: | Silicone | Application: | Clinic |
High Light: | disposable insulin syringe,disposable scalp vein set |
Hemodialysis Catheter Infusion And Injection Medical Grade PVC For Clinic
Configuration:
1. Dilator
2. Scalpel
3. Introdocer Needle
4. Springe-shape introducer
5. Hemodialysis catheter
6. Guidewire with advancer
7. Injection sit
8. Dilator
5CM | 8CM | 13CM | 16CM | 20CM | |
5.5Fr |
Infant Jugular |
Infant Subclavia |
Infant Femoral |
||
6.5Fr |
Child Jugular |
Infant Subclavia |
Infant Femoral |
||
8Fr |
Teenage Jugular |
Teenage Subclavia |
Teenage Femoral |
||
11.5Fr |
Adult Jugular |
Adult Subclavia |
Adult Femoral |
||
12Fr |
Adult Jugular |
Adult Subclavia |
Adult Femoral |
Catheterization procedure
1. Place the patient in a trendelenberg position (15 to 30 degrees head down) to reduce the chance of an air embolism, turn the patient’s head to the side contralateral to the site chosen.
2. mark puncture site
3. Anesthetize the spot by using needle, the angle between needle and skin shall be 30 to 45 degrees.
4. when a flush of blood returns by advancing the needle, insert venous needle.
Remarks: pay attention to the direction of needle and color of blood returns during advancing seeker needle , meanwhile, pay attention to the direction of needle and depth of puncturing needle , the bevel of needle shall be faceup.
5. one hand hold the needle,another hand hold the syringe ,keep the pressure negative during the slow insertion. When you feel the needle placed into the vein, the blood will return smoothly.
6. Guidewire will be inserted through Y shaped syringe or hollow syringe.
Remarks: during the insertion of guidewire,please judge the resistence , don’t withdraw
the guidewire by force, withdraw the guidewire together with syringe if ncessary.
7. withdraw syringe, and insert dilator by 1-2 cm under the skin for enlarge the track.
Remarks: During insertion of dilator, attention shall be paid to the direction and the force of dilator, specially the force transfering between dilator and guidewire, the driving force shall be increased gradually
Withdraw the dilator, implant the catheter from the end of guidewire, and saline shall be injected into to the catheter immediately.
Ensure the whole catheter implanted completely and then fixed the catheter at the final position.
Connect with other tubes and devices
The force for inserting the catheter is related to enlarging of the track, if the resistence is big, the length of each insertion shall be shorter , and attetion shall be paid to avoid kink of guidewire.